Self-registration of Ultrasound Imaging Device to Navigation System Using Surgical Instrument Kinematics in Minimally Invasive Procedure
- 章节作者
- 章节序号
-
8
- 章节页码
-
95-103
- 著作名称
-
Computer Aided Surgery
- 图书作者
- 出版社
-
Springer
- P-ISBN
-
9784431558088
- E-ISBN
-
978-4-431-55810-1
- 摘要
-
This work presents an intraoperative method for registration of 3D ultrasound images to surgical navigation systems. Unlike conventional methods that rely on pre-calibration of ultrasound probe in a laboratory environment using a calibration rig, the proposed method is designed to self-register with a tracking system in an operating room. It exploits intraoperative motion of the surgical instrument to compute the registration function instead of using the known geometry of a phantom structure. To reduce estimation errors arising from poor image quality and uncertainty of image-based position estimations, the pivotal constraint associated with minimally invasive surgeries is imposed as a positional correspondence between the imaging and tracking system. We demonstrated the registration of an ultrasound imaging device to an optical tracker and evaluated the accuracy in a water tank setup. This self-registration framework has potential contribution in the unification of ultrasound image-guidance and tracker-based surgical navigation systems. The ultimate goal is to combine ultrasound image-guidance and tracking systems as an integrated surgical navigation system.
- DOI
- https://doi.org/10.1007/978-4-431-55810-1_8